Performance and operational feasibility of antigen and antibody rapid diagnostic tests for COVID-19 in symptomatic and asymptomatic patients in Cameroon: a clinical, prospective, diagnostic accuracy study

https://doi.org/10.1016/S1473-3099(21)00132-8Get rights and content

Summary

Background

Real-time PCR is recommended to detect SARS-CoV-2 infection. However, PCR availability is restricted in most countries. Rapid diagnostic tests are considered acceptable alternatives, but data are lacking on their performance. We assessed the performance of four antibody-based rapid diagnostic tests and one antigen-based rapid diagnostic test for detecting SARS-CoV-2 infection in the community in Cameroon.

Methods

In this clinical, prospective, diagnostic accuracy study, we enrolled individuals aged at least 21 years who were either symptomatic and suspected of having COVID-19 or asymptomatic and presented for screening. We tested peripheral blood for SARS-CoV-2 antibodies using the Innovita (Biological Technology; Beijing, China), Wondfo (Guangzhou Wondfo Biotech; Guangzhou, China), SD Biosensor (SD Biosensor; Gyeonggi-do, South Korea), and Runkun tests (Runkun Pharmaceutical; Hunan, China), and nasopharyngeal swabs for SARS-CoV-2 antigen using the SD Biosensor test. Antigen rapid diagnostic tests were compared with Abbott PCR testing (Abbott; Abbott Park, IL, USA), and antibody rapid diagnostic tests were compared with Biomerieux immunoassays (Biomerieux; Marcy l'Etoile, France). We retrospectively tested two diagnostic algorithms that incorporated rapid diagnostic tests for symptomatic and asymptomatic patients using simulation modelling.

Findings

1195 participants were enrolled in the study. 347 (29%) tested SARS-CoV-2 PCR-positive, 223 (19%) rapid diagnostic test antigen-positive, and 478 (40%) rapid diagnostic test antibody-positive. Antigen-based rapid diagnostic test sensitivity was 80·0% (95% CI 71·0–88·0) in the first 7 days after symptom onset, but antibody-based rapid diagnostic tests had only 26·8% sensitivity (18·3–36·8). Antibody rapid diagnostic test sensitivity increased to 76·4% (70·1–82·0) 14 days after symptom onset. Among asymptomatic participants, the sensitivity of antigen-based and antibody-based rapid diagnostic tests were 37·0% (27·0–48·0) and 50·7% (42·2–59·1), respectively. Cohen's κ showed substantial agreement between Wondfo antibody rapid diagnostic test and gold-standard ELISA (κ=0·76; sensitivity 0·98) and between Biosensor and ELISA (κ=0·60; sensitivity 0·94). Innovita (κ=0·47; sensitivity 0·93) and Runkun (κ=0·43; sensitivity 0·76) showed moderate agreement. An antigen-based retrospective algorithm applied to symptomatic patients showed 94·0% sensitivity and 91·0% specificity in the first 7 days after symptom onset. For asymptomatic participants, the algorithm showed a sensitivity of 34% (95% CI 23·0–44·0) and a specificity of 92·0% (88·0–96·0).

Interpretation

Rapid diagnostic tests had good overall sensitivity for diagnosing SARS-CoV-2 infection. Rapid diagnostic tests could be incorporated into efficient testing algorithms as an alternative to PCR to decrease diagnostic delays and onward viral transmission.

Funding

Médecins Sans Frontières WACA and Médecins Sans Frontières OCG.