Diagnostic accuracy of LAMP versus PCR over the course of SARS-CoV-2 infection

Int J Infect Dis. 2021 Jun:107:195-200. doi: 10.1016/j.ijid.2021.04.018. Epub 2021 Apr 20.

Abstract

Objective: Reverse transcription loop-mediated isothermal amplification (RT-LAMP) has been validated to diagnose several viral infections. However, its diagnostic accuracy in detecting SARS-CoV-2 in real-life clinical settings remains unclear. This study aimed to determine the diagnostic sensitivity and specificity of RT-LAMP compared to reverse transcription-quantitative polymerase chain reaction (RT-qPCR) over the disease course of COVID-19.

Methods: A total of 124 nasopharyngeal swab samples obtained from 24 COVID-19 patients were tested by RT-LAMP and RT-qPCR. Sensitivities and specificities of RT-LAMP compared with RT-qPCR were analyzed as a function of time from onset.

Results: Up to the 9th day after onset, the RT-LAMP had a positivity of 92.8%, and the sensitivity and specificity compared with RT-qPCR was 100%. However, after the 10th day after onset, the positivity of RT-LAMP decreased to less than 25%, and the concordance of positivity between the two methods was below 60%. The limit of detection of RT-LAMP was 6.7 copies/reaction.

Conclusions: Until the 9th day after the onset of symptoms, RT-LAMP had the same diagnostic accuracy as RT-qPCR. These findings suggest that RT-LAMP can be used as a diagnostic tool for COVID-19 as an alternative to RT-qPCR in the acute symptomatic phase of COVID-19.

Keywords: COVID-19; RT-LAMP; RT-qPCR; Sensitivity; Specificity.

MeSH terms

  • Adult
  • COVID-19 / diagnosis*
  • COVID-19 Nucleic Acid Testing / methods*
  • COVID-19 Testing*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Molecular Diagnostic Techniques / methods*
  • Nucleic Acid Amplification Techniques / methods*
  • SARS-CoV-2*
  • Sensitivity and Specificity

Supplementary concepts

  • LAMP assay