Performance characteristics of an automated latex immunoturbidimetric assay [HemosIL® HIT-Ab(PF4-H)] for the diagnosis of immune heparin-induced thrombocytopenia

Thromb Res. 2017 May:153:108-117. doi: 10.1016/j.thromres.2017.03.010. Epub 2017 Mar 11.

Abstract

Background: Heparin-induced thrombocytopenia (HIT) is a prothrombotic drug reaction caused by platelet-activating anti-PF4/heparin antibodies. Given time-sensitive treatment considerations, a rapid and accurate laboratory test for HIT antibodies is needed.

Aims: To determine operating characteristics for the HemosIL® HIT-Ab(PF4/H), a rapid, on-demand, fully-automated, latex immunoturbidimetric assay (LIA), for diagnosis of HIT.

Methods: We evaluated LIA sensitivity, specificity, negative (NPV) and positive predictive value (PPV), negative (LR-) and positive likelihood ratio (LR+), using citrated-plasma from 429 patients (prospective cohort study of 4Ts scoring; HIT, n=31), and from consecutive HIT patients (n=125), using reference standard serotonin-release assay (SRA). Comparators included two PF4-dependent enzyme-immunoassays (EIAs). We used stratum-specific likelihood ratios (SSLRs) to determine how differing magnitudes of LIA-positivity influenced post-test probability of HIT.

Results: LIA operating characteristics were: sensitivity=97.4% (152/156); specificity=94.0% (374/398); PPV=55.6% (30/54); and NPV=99.7% (374/375). At manufacturers' cutoffs, LIA specificity and PPV were superior to the EIAs. Although a negative LIA pointed strongly against HIT (LR-, 0.034), the post-test probability was ~2% with high 4Ts score. The LIA's LR+ was high (16.0), with SSLRs rising substantially with greater LIA-positivity: 5.7 (1.0-4.9U/mL), 31 (5.0-15.9U/mL), and 128 (≥16U/mL). A LIA-positive result (at 1.0 cutoff) indicated at least 24% HIT probability (low 4Ts score), rising to 90% with high 4Ts score.

Conclusions: Although approximately 1 in 40 SRA-positive patients tested LIA-negative, the LIA's high NPV and PPV indicate that this rapid assay is useful for the diagnostic evaluation of HIT, including in low pre-test situations.

Keywords: Enzyme-immunoassay; Heparin-induced thrombocytopenia; Latex immunoturbidimetric assay; Rapid assay.

Publication types

  • Evaluation Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Antibodies / immunology
  • Anticoagulants / adverse effects*
  • Anticoagulants / immunology
  • Female
  • Heparin / adverse effects*
  • Heparin / immunology
  • Humans
  • Immunoturbidimetry / economics
  • Immunoturbidimetry / methods*
  • Latex / immunology
  • Male
  • Middle Aged
  • Platelet Factor 4 / immunology
  • Prospective Studies
  • Purpura, Thrombocytopenic, Idiopathic / chemically induced*
  • Purpura, Thrombocytopenic, Idiopathic / diagnosis*
  • Purpura, Thrombocytopenic, Idiopathic / immunology
  • Sensitivity and Specificity

Substances

  • Antibodies
  • Anticoagulants
  • Latex
  • Platelet Factor 4
  • Heparin