Highly Stable Hierarchically Structured All-Polymeric Lubricant-Infused Films Prevent Thrombosis and Repel Multidrug-Resistant Pathogens

ACS Appl Mater Interfaces. 2022 Dec 7;14(48):53535-53545. doi: 10.1021/acsami.2c17309. Epub 2022 Nov 22.

Abstract

Thrombus formation and infections caused by bacterial adhesion are the most common causes of failure in blood-contacting medical devices. Reducing the interaction of pathogens using repellent surfaces has proven to be a successful strategy in preventing device failure. However, designing scale-up methodologies to create large-scale repellent surfaces remains challenging. To address this need, we have created an all-polymeric lubricant-infused system using an industrially viable swelling-coagulation solvent (S-C) method. This induces hierarchically structured micro/nano features onto the surface, enabling improved lubricant infusion. Poly(3,3,3-trifluoropropylmethylsiloxane) (PTFS) was used as the lubricant of choice, a previously unexplored omniphobic nonvolatile silicone oil. This resulted in all-polymeric liquid-infused surfaces that are transparent and flexible with long-term stability. Repellent properties have been demonstrated using human whole blood and methicillin-resistant Staphylococcus aureus (MRSA) bacteria matrices, with lubricated surfaces showing 93% reduction in blood stains and 96.7% reduction in bacterial adherence. The developed material has the potential to prevent blood and pathogenic contamination for a range biomedical devices within healthcare settings.

Keywords: antibacterial; antithrombogenic; hierarchically structured surfaces; liquid-infused surfaces; omniphobic lubricant.

MeSH terms

  • Blood Stains*
  • Humans
  • Lubricants / pharmacology
  • Methicillin-Resistant Staphylococcus aureus*

Substances

  • Lubricants