Defining humeral axial rotation with optical motion capture and inertial measurement units during functional task assessment

Med Biol Eng Comput. 2023 Nov;61(11):2963-2970. doi: 10.1007/s11517-023-02894-z. Epub 2023 Aug 3.

Abstract

Humeral motion can be challenging to measure and analyze. Typically, Euler/Cardan sequences are used for humeral angle decomposition, but choice of rotation sequence has substantial effects on outcomes. A new method called True axial rotation calculation may be more precise. The objective of this study is to compare humeral axial rotation measured from two systems (optical motion capture and inertial measurement units (IMUs)) and calculated with two methods (Euler angles and True axial). Motion of torso and dominant humerus of thirty participants free from any upper limb impairments was tracked using both systems. Each participant performed a functional tasks protocol. Humeral axial rotation was calculated with Euler decomposition and the True axial method. Waveforms were compared with two-way ANOVA statistical parametric mapping. A consistent pattern emerged: axial rotation was not different between motion capture systems when using the True axial method (p > .05), but motion capture systems showed relatively large magnitude differences (~ 20-30°) when using Euler angle calculation. Between-calculation method differences were large for both motion capture systems. Findings suggest that the True axial rotation method may result in more consistent findings that will allow for precise measurements and comparison between motion capture systems. Two methods for calculating humeral axial rotation measured from optical motion capture and inertial measurement units were compared.

Keywords: 3D kinematics; Euler angles; Functional; Humeral; Shoulder.

MeSH terms

  • Biomechanical Phenomena
  • Humans
  • Humerus
  • Motion Capture
  • Range of Motion, Articular
  • Rotation
  • Shoulder Joint*