Evaluation of the Effect of Gait Aids, Such as Canes, Crutches, and Walkers, on the Accuracy of Step Counters in Healthy Individuals

Orthop Res Rev. 2021 Jan 7:13:1-8. doi: 10.2147/ORR.S292255. eCollection 2021.

Abstract

Background: Wearable step-counter devices have become inexpensive tools that enable patients, researchers, and clinicians to objectively monitor physical activity. It is unknown how the use of gait aids, such as canes, crutches, and walkers, affects the accuracy of these devices. Such gait aids are commonly used by patients with chronic physical impairment and after joint-replacement surgery. The aim of this study was to determine the effect of gait aids on the accuracy of wearable step counters.

Methods: Nine healthy participants wore a Fitbit step counter on their wrist and hip and performed eight walking tests with canes, crutches, and walkers. Bland-Altman analyses were performed for all eight walking tests in order to compare agreement between measurement techniques.

Results: Mean overall agreement for subjects walking without gait aids in the hip group was excellent, showing a bias of -2.9, with limits of agreement (LOAs) between -8.72 and 2.95. For use of canes or crutches, the Bland-Altman plots had a range of bias values between the hip and wrist counters from -7.22 to 33.56, with LOAs from -98.55 to 124.2. The wrist counter during the four-wheeled walking test showed very little agreement with the actual step count, with a bias value of 91.33 and LOAs of 64.1-118.6, illustrating exceptionally unreliable step counts.

Conclusion: This study suggests that these widely commercially available step counters have poor reliability with gait aids, especially walkers, which should be taken into account in research and clinical settings.

Keywords: Fitbit; gait aids; monitor; physical activity; rehabilitation; step counters.